How To Solve The Problems Of Constant Temperature And Humidity Test Chamber?
When constant temperature and humidity test chamber have some problems during using,what we can do to solve it?
1. Water Leakage
Solution: Install pressure balance port to ensure consistent pressure inside and outside the box and solve the problem of water leakage
2. High Pressure Alarm
A: the test environment temperature of the air-cooled constant temperature and humidity is too high
Solution 1: It is recommended that the customer's ambient temperature<30°C (usually controlled at 20-27°C)
B: the circulation fan is faulty, the circuit board capacitor that controls the fan is faulty, causing the speed to not meet the standard, and the temperature is too high, resulting in an alarm
Solution 2: Change the circulating fan
3. The Entrance Of The Painted Water Tank Is Easy To Rust
Solution: Communicate with the paint shell supplier to improve the stainless steel treatment
4. Compressor Noise
Solution: The bottom is not sealed, only a fixed frame or customer non-standard requirements, increase the thickness of the plate
5. Evaporator Is Frosted
Solution: An additional evaporating pressure regulating valve is required to increase the evaporating pressure on the surface of the evaporator, so as to control the evaporating temperature and prevent frosting
If some other problems or you cannot solve these issues,please contact us to help you.